The ROU28S is a receiver designed for MASFLO-OVAL Ⅱ (thermal mass flowmeter/controller).
Functions necessary for flow measurement/control are integrated into specialized units.
The ROU28S enables the simple and cost-effective configuration of flow measurement/control systems.
●Small and lightweight receiver equipped with basic functions necessary for flow measurement/control.
●Flow rate value, accumulated flowrate value and flowrate set-point can be switched to the display.
●A flow rate adjustment device and an external input switch are placed on the front, realizing an easy-to-use layout.
●As standard specifications, it includes external output (analog and pulse) and external control input functions.
Therefore, flexibility of system design can be improved.
●By combining the readout unit with the MASFLO-OVAL II series, a flow measurement system can be easily configured.
Item | Description | ||
---|---|---|---|
Measurement | Measurement types | Instantaneous flowrate value/accumulated flowrate value | |
Measurement method | A/D conversion method Resolution: Approximately 1/7000 (relative to full-scale input) |
||
Input measurement interval: Approximately 20 ms | |||
Low-cut | 0 to 29% of maximum analog input (can be set arbitrarily) | ||
Display | Display | Red LED, Character Height: 8 mm (Zero-Blanking Method) | |
Display switching | Switch between displaying flowrate value and accumulated value using the ▲ key (Requires setting in “Mode No.4”) ※When the DISP switch is set to OUT |
||
RA lamp | Lights up (red) during flowrate value display | ||
TO lamp | Lights up (red) during accumulated value display | ||
OV lamp | Lights up (red) when the displayed measurement value exceeds the range (During right-shifted display, it blinks) |
||
OUT1 lamp | Lights up during output of Alarm output OUT1 | ||
OUT2 lamp | Lights up during output of Alarm output OUT2 | ||
Instantaneous value display |
Measuring accuracy | ±0.3% (F.S.) ±1 digit (at 23°C) for analog input | |
Scaling method | Arbitrary setting of instantaneous flowrate value display at maximum analog input (0.001 to 9999) |
||
Display range | 0 to 9999 (When the displayed value exceeds this range, “9999” blinks and OV Lamp lights up) | ||
Decimal point representation | Set at the maximum display value (Mode No.1). | ||
Display sampling time | Averaging the display in approximately 0.5 seconds (fixed) | ||
Least significant digit correction |
Averaging the display in approximately 0.5 seconds (fixed)
Least Significant Digit Correction: Selectable from “Normal,” “0 fixed,” and “0 or 5” |
||
Accumulated value display |
Measuring accuracy | ±0.3% (F.S.) ±1 digit (at 23°C) for analog input | |
Scaling method | Arbitrary setting of accumulated value per hour at the maximum analog input value by mode setting | ||
Display range | 0 to 99999 | ||
Decimal point display | Selectable from 1 to 4 decimal places. | ||
Over (OV) display | When the displayed value exceeds 99999, it automatically shifts one digit to the right (continuing with right-shifted display). During right-shifted display, the OV Lamp blinks. If the display exceeds the x10 scale, the OV Lamp lights up and the display becomes endless. |
||
Display offset | The accumulated value after reset can be arbitrarily set in the range of 0 to 99999 by setting the display offset value. | ||
Input signal | Flowrate signal | DC 0 to 5V, D-sub, input resistance 330KΩ | |
External setting signal | DC 0 to 5V, terminal block, input resistance 330 KΩ | ||
VOR IN | Not used for MASFLO-OVAL II | ||
Operating section | Mechanical dial (CMDSET) |
Minimum scale of 10 rotary type potentiometer: 1/50 turns Equipped with a lock lever (to fix rotation and scale) (The scale on the dial serves as a reference for the amount of rotation) |
|
Set it to the desired flowrate for control | |||
VOR switch | Not used for MASFLO-OVAL II (CONT only) | ||
Command changeover switch (COMMAND) |
EXT: The external setting value signal becomes the display for the flowrate setpoint. INT: The CMDSET mechanical dial (CMD) setting becomes the flowrate set-point display. |
||
Display changeover switch (DISP) |
SET: Displays the flowrate set-point (relative to COMMAND). OUT: Displays the instantaneous and accumulated flowrate values. |
||
Operation setting switch | Key switches of MODE, ▲, and RST | ||
Output-related | Power supply rating | ±15VDC, 250mA each or DC +15V 150mA and DC -15V 350mA also supported Ripple noise: 30mV or less |
|
Setting signal output | DC 0 to 5V, 5mA, D-sub | ||
External signal output | DC 0 to 5V, terminal block | ||
Valve opening/closing output | Not used for MASFLO-OVAL II | ||
Alarm and synchronous output |
Output terminal | Output from terminal blocks 11-12 (OUT1) and 13-12 (OUT2) (common to COM) | |
Comparison method | Selectable from upper and lower limits | ||
Output mode | Comparison/synchronization pulse (OUT1: Selectable via mode) | ||
Preset value setting | Can be set arbitrarily in the preset value setting mode. | ||
Alarm, synchronous output, and others |
Output judgment | Output determined by comparing the displayed value with the preset value | |
Output method | NPN open collector output, 2 points
Maximum rating: DC30V, 50mA, MAX |
||
Data backup | Writes the setting values for each mode and accumulated measurement to FRAM (Rewriting limit: 100,000 times or less,
approximately 10 years of retention) |
||
Mode protect function | Switching selection based on mode protect setting
When “ON,” the set value cannot be changed; when “OFF,” the set value can be changed |
||
Warm-up time | 30 minutes or more after power-on | ||
Power supply | AC 100 to 240V (-15% / +10%), 270 mA max
50/60Hz, approx. 35VA or less |
||
Others | Operating temperature and humidity ranges |
0 to 50°C, 30 to 80%RH (without condensation) | |
Weight and outside dimensions | Approximately 400g, W96 × H48 × D156 mm | ||
Main body color | Black | ||
Housing material | ABS | ||
Operating environment | Indoor use, Maximum altitude: 2,000m, Overvoltage category II, Pollution level 2 | ||
Low voltage directive | EN61010-1 | ||
EMC | EN61326-1 EN55011(Group1 ClassA)、EN61000-4-2、EN61000-4-3、 EN61000-4-4、EN61000-4-5、EN61000-4-6、EN61000-4-11 |
*You need to fill out the questionnaire to download documents.
ROU28S