//=$_SERVER["REQUEST_URI"] == SRC_ROOT ? 'website': 'article'?>
PRODUCTS
Format:EL4111
Using the most advanced electronic technologies, this digital instrument has been developed specifically to meet the most demanding steam flow measurement applications where accuracy is the prime requirement.
In response to flowrate, temperature, and pressure information arriving from the sensing terminal, such as a Vortex flowmeter, it calculates superheated steam flow, transforms it to mass and calorie, and totalizes the flow.
It also provides a totalizer pulse output and an instantaneous flowrate analog output(mass equivalent).
1. By simple key operations or by inserting an IC card in the slot, parameters such as meter factor, temperature and pressure ranges and other parameters in the connected flowmeter can be easily changed.
2. Built around a microprocessor, the instrument carries out calculations entirely in digital signal processing circuits to achieve high degree of accuracy and reliability.
3. Variables such as temperature and pressure can be reviewed on command with the front panel keypad even during operation.
4. A nonvolatile memory (E 2 PROM) retains all parameters and variables. Totalized value can be reset upon power depending on setting.
Kinds | STEAM FLOW COMPUTER (SUPERHEATED STEAM SERVICE) | ||
---|---|---|---|
Type | EL4111 (Superheated Steam Service) | ||
Power supply for combined flowmeter | 13.5 or 24 VDC (depends on combined flowmeter) | ||
Display | Total flow (mass, calorie), instantaneous mass flowrate, specific weight, pressure, temperature, other | ||
Input signal | Flowrate (pulse), pressure, temperature | ||
Function | Scaler, frequency division setting, F/I conversion, instrumental error correction, pressure correction, temperature correction, communication | ||
Output signal | Analog, pulse, alarm | ||
Computation range | Temperature | 100 to 350°C (Span limited) | |
Pressure | 0 to 3 MPa, standard span: 1 MPa | ||
Computing accuracy | Mass flow | ±0.5% RD | |
Calorie flow | ±0.6% RD | ||
Temperature | ±0.1% of SPAN or ±0.3% of SPAN | ||
Pressure | ±0.1% of SPAN | ||
Ambient temperature | –10 to +50°C | ||
Construction | Non-explosionproof | ||
Power supply | AC- or DC-powered |
*You need to fill out the questionnaire to download documents.