Flowmeters and System Solutions

Flowmeters and System solution.

MENU

PRODUCTS

DENSITY COMPUTER FOR MASS FLOWMETER

Format:EL5300

This digital instrument has been developed to meet the demands for high-precision fluid measurement.

This calculator (or "the instrument") receives tube vibration signals and temperature signals from an oval Coriolis flowmeter, then calculates and displays the fluid's density. It also displays the concentration of solid components in slurry
as a percentage. Furthermore, it is capable of displaying and analog outputting
the corrected density converted to a reference temperature.

Features
Product Specifications

FEATURES

1. Equipped with a large color LCD touch panel, the instrument allows for intuitive parameter setting and operation. It also features a customizable display screen, which helps prevent reading errors by allowing users to hide items not required for inspection.

2. Supporting Modbus communication, the instrument enables reading of various measured values such as integrated values and event logs, as well as writing of various parameters.

3. Various parameters, event logs, and other data stored internally can be easily extracted to a USB memory stick.

GENERAL SPECIFICATION

Density After Correction・ULTRAmass MKⅡ (CN series)
±0.002g/mL (when temperature input is 4 to 20mADC, 1 to 5VDC),
±0.006g/mL (when temperature input is Pt100Ω at 0℃ 3-wire type)※1

Kinds DENSITY COMPUTER FOR MASS FLOWMETER
Type EL5300
Display Density (Before Correction / After Correction), Temperature, etc.
Input Signal Temperature, Density
Output Signal Analog Output, Alarm Output
Calculation Range Temperature Resistance Thermometer (Pt100Ω) / Range: −50 to +200℃
Analog (4 to 20mADC / 1 to 5VDC) / Range: −20 to +75℃ Standard Span: 200℃
Analog (10mV/℃) / Range: −50 to +150℃ Standard Span: 200℃
Analog (5mV/℃) / Range: −200 to +200℃ Standard Span: 200℃
Density 0.0000 to 6.0000g/mL
Calculation Accuracy Density Before Correction ・ULTRAmass MKⅡ (CN series)
±0.001g/mL (Factory Calibration Accuracy: Optional)
Temperature Resistance Thermometer (Pt100Ω): ±0.2℃
Analog (4 to 20mADC / 1 to 5VDC): ±0.1% of SPAN
Analog (10mV/℃): ±0.1% of SPAN
Analog (5mV/℃): ±0.1% of SPAN
Ambient Temperature −10 to +50℃
Power Consumption 20W Max
Power Supply AC or DC Power Supply

※1: Depends on the accuracy of the temperature input and the density expansion coefficient of the fluid relative to the density before correction.
(This shows the accuracy when the density change is constant relative to the difference between the temperature input and the reference temperature.)